MHJ -1C Anesthesia Machine
It adopts a two-stage pressure regulator and a high precision dual-tube flowmeter to provide an accurate fresh gas
MHJ -1C Anesthesia Machine
It adopts a two-stage pressure regulator and a high precision dual-tube flowmeter to provide an accurate fresh gas output.Its build-in safety devices such as ORC system, oxygen deficiency alarm and N20 cut-off make anaesthesia more safe. The specially constructed inter-lock mechanism may guarantee only one of the two parallel vaporizers to be switched on at a time.The selected vaporizers from world famous companies as Drager 19.3,Ohmeda TEC-5 and Penlon Sigma Delta satisfy different requirements of customers.With MHJ -1C it is possible to act low-flow anaesthesia because of its unique integrated circle system-a system that keeps the inhaled tidal volume from fresh gas. So it not only saves anaesthesia agent but also reduces pollution to operation room.
Main Specifications:
Flow:O2 0.1~ 15L /min ; N2O 0.05~ 10L /min ; ORC O2 = 25% O2 Flush: 35~ 75L /min
Oxygen deficiency alarm and N2O cut-off :when O2 = 0.2Mpa,audible alarm occurs and lasts for at least 7 seconds ; when O2 = 0.14Mpa,N2O flow decreases correspondingly ; when O2 = 0.04Mpa,N2O flow is cut off Integrated breathing circle system: APL adjustment range:0.5~7kPa ; CO2 absorber acpcity: 1.6L
Vaporizer:
Ambient temperature 15~ 35 ?
Input flow 0.25~ 15L /min
Output back pressure:-10kPa~20kPa
Output concentration:0.2~5Vol%±20%(OR±0.2%)
Anesthesia Ventilator:
A microprocessor-controlled ventilator,which is time-cycled and constant-volume,has a clear digital reading for all set breathing parameters.Because it is electric-driven and electric-controlled,the ventilator saves large amount of oxygen that is usually used only as driving gas.
Specifications:
Frequency:6-60bmp ;Tidal volume:100~1100mL ; I:E ratio 1:1~1:9.9 adjustable
Monitor:
Respiratory monitor applies microprocessor to monitor Tidal Volume,Frequency,MV airway pressure,and inhaled oxygen concentration.It has real time curve display and a visual-and-audible alarms.A Chinese menu and single-key operation is userfriendly. A high-definition large screen with wide vision angle make all-directional inspection available.
Specification:Frequency:6-60bmp
O2 measuring range 0~99%
Tidal Volume:100~1500mL
Minute volume:0 -40L /min
Airway pressuer:0-9.9kPa